Research
|
||||||||||||||||||||||||||||||||||||
|
Medical and biomedical image analysisDigital mammographyOUTLINE
Breast cancer is still the most common form of cancer among women.
Mammography (Fig. 1) is a specific type of imaging that uses a low-dose x-ray system
to examine breasts.
Digital mammography, often called Full-Field Digital Mammography (FFDM) [MI1],
is a mammography system in which the x-ray film is replaced by solid-state
detectors that convert x-rays into electrical signals. These detectors are similar
to those found in digital cameras.
Mammography is currently the best imaging technique for early detection and
diagnosis of breast cancers in women because it can show changes in the breast
up to two years before a patient or a doctor can feel them. Fig. 1 also shows different kinds of breast cancers,
microcalcifications and masses [MI2].
Microcalcifications are very small spots that are relatively bright compared with the surrounding normal tissue.
Typically they are between 0.1 mm and 1.0 mm in size and their presence in breast tissues is
of particular clinical significance when found in cluster of five of more in 1 cm2 area.
Most of the clusters consist of at least one evident microcalcification and other more hidden signals.
Fig. 1: From left to right: Cranio-caudal (CC) and mediolateral-oblique (MLO) views in mammograms; examples of microcalcifications and mass. CAD (computer-aided detect) systems use a digitally acquired mammogram to assist radiologists to detect breast cancers: the computer output is presented to radiologists as a second opinion and can improve the accuracy of the detection. The CAD system highlights these regions on the images, alerting the radiologist to the need for further analysis.Several techniques developed for the automated detection of microcalcifications can mainly be grouped into three different categories: multiresolution analysis, difference-image techniques and statistical methods. By comparing the different methods it turns out that some microcalcifications are detected by one method but missed by others, due to different types of microcalcifications. It is thus often hard for one single detection scheme to discover different types of signal with diverse characteristics. METHOD The method we propose combines the coarse- and fine-level detection methods we developed in order to get optimal performance [MI3][MI4]. The CAD system we realized is able to perform automatic detection of microcalcifications. Fig. 2 shows the block-based flow chart of our system. Both the methods are able to find out the most evident signals, i.e. having high local contrast and high Signal to Noise Ratio (SNR). Fig. 2: The block-based functional scheme of our CAD system. Fig. 3: Coarse level signal extraction. The original images are 12 bits grey levels with a resolution of 0.1 mm/pixel. In the coarse level detection, two different operations are performed starting from this image, contrast enhancement and signal suppression. In particular, the signal suppression is achieved through decomposing the image using wavelet transform (where the mother wavelet is LAD8) and reconstructing the signal by keeping low frequencies only. Also, this means extracting the structured background from the image. Then the difference between these images ends the preprocessing step. Basically, this represents a background difference and the same principle (although not the same technique) has been employed even in our motion detection algorithms. This image is now supposed to contain only signals in the high grey level domain. This means that the Gaussian distribution of the grey levels corresponding to the background should be biased towards the high levels, thus expecting a positive skewness. For these reason a local thresholding function of pixel grey levels and skewness is performed to achieve a binary image of the microcalcifications (Fig. 3).The fine level detection has been conceived to discover more subtle microcalcifications, by means of a multiresolution analysis based on the wavelet transform (LAD8). To reconstruct the image we use the first three scales, corresponding respectively to details of 100µm, 200µm and 400µm. Although scale 1 retains high frequency noise, it also retains useful information regarding microcalcifications. Therefore, rather than discard it we work by keeping only the coefficients in scale 1 greater that 1.5σ, referring to the coefficient distributions. To extract interesting signals, we perform in the resconstructed image a local thresholding in a 40×40 window retaining all pixels having grey level values above 2.0σ. The same thresholding is performed on the image achieved through the difference between a suppressed image (obtained with a 9×9 moving average filter) and an enhanced image (coming from a convolution with a 3×3 match filter). The output of the fine level detection is the binary image attained by AND'ing the output of these thresholdings. To summarize the performances of our two signal extraction methods, we can state that the coarse one is able to find out even microcalcifications having a low contrast, provided that they retain a high SNR. Nevertheless, the fine level detection is suitable to find out even low SNR signals, provided that they have a high contrast, thanks to the wavelet transform that can enhance in a better way signals with similar size as microcalcifications. Then the output of both the methods is OR'ed and the binary image is then used as a mask to identify the microcalcifications in the original image. More details are reported in [MI3][MI4]. False Positive Reduction (FPR). Most of the signals survived to the thresholding stages are False Positive
ones that have to be eliminated. A thorough discussion with radiologists has permitted to recover the
method they use to detect microcalcifications and to turn it into digital features that can be computed
automatically. Accordingly, we have considered five features (area, average pixel value, edge gradient,
degree of linearity, average local gradient) to separate microcalcifications from false signals [MI4][MI5].
Therefore for each labelled (true, false) signal has been provided with this five-dimensional feature
vector that constitute the input of a Support Vector Machines (SVM) classifier.
Clustering. Simply we define a cluster as being an ensemble of at least three microcalcifications whose minimum distance between them is no more than 0.5 mm. RESULTS
Here we show a comparison among the performances of the three classifiers to reduce false signals.
We have randomly divided the 9000 detected signals into 3 groups for 9 different times and
the results have been averaged.
Fig. 4: Performance of our CAD system after FPR reduction (left); the ROC curves of the three classifiers with full size (middl) and reduced size (right) training sets. Fig. 4 (left) illustrates the overall quality performance of our CAD system using SVM, through the FROC (Free Response Operating Characteristic) curve. Our CAD yields a sensitivity of 95% true clusters with 0.6 false-positive clusters on the 40 images of the Nijmegen database. The output is computed as the average on the whole database of the SVM classifier trained on the 9 training groups. Fig. 4 (middle) shows the ROC curves of the three classifiers, where LDA has the worst performance. Although SVM and MLP shows comparable performances, the SVM classifier is far simpler to handle and setup. These results are comparable with the state-of-the art for this database, the reference one in this field. Finally, we test the robustness of the classifiers when trained with a reduced training set. Fig. 4 (right) shows the variation of Az as a function of the size of the training set. Here SVM outperforms the MLP classifier, yet more when the training set is small.Fig. 5 shows the simple Graphical User Interface we developed to prompt the radiologist in case of detection of a cluster of microcalcifications. Fig. 5: The original image (left), with the highlighted cluster found out by our CAD (middle) and the cluster at half the original size (right). It is worth remarking that we have patented our CAD system that has been included in a wider CAD system, now used in several Hospitals in the world.3D Positron Emission Tomography(3D PET)OUTLINE
Positron Emission Tomography (PET) is a tomographic method which allows imaging of any body part
of a patient to whom has been previously administered a radiopharmaceutical. The radioactive isotopes
(tracers) are targeted to specific molecules, organs or tissues by attaching them to other compounds
such as glucose. The image is created by measuring this distribution inside the body. In particular,
a positron resulting from isotope decay annihilates with an electron in the surrounding body tissue, producing
two 511 keV photons (events) travelling in opposite direction. The coincidence detection of these
opposite photons allows an image of the density of the annihilation points to be reconstructed by approximating
the spatial distribution of the tracer.
METHOD
The method we conceived performs a fully 3D image reconstruction, where also photons with trajectories
oblique to the plane of detectors are considered, thus increasing significantly the sensitivity of the system.
The method has been implemented in a parallel distributed algorithm to work in real time on a distributed
computer made of a Cluster of Workstations.
We have conceived a BPF method for a fully 3D image reconstruction because it works on list-mode data,
thus being suitable to work on-line. More details are given in [M12].
Fig. 1: The scheme of a classical PET scanner (left) and our experimental PET apparatus for small animals (right). Fig. 2: Left: Backprojection scheme in a 2D section of the image matrix: the voxels intersected (shaded) by the LORs (dot lines) are filling with weighted values. Right: A pair of opposite detectors with geometrical centre O. Each LOR is subdivided into equally-spaced points; through each of them (e.g.: P) a number of LORs could pass which is geometrically limited by the maximum number of crystals that could be fired. Here, if an event had happened at point P, then the maximum number of LORs that could have been detected theoretically would be 25 (left shaded surface). Each LOR is subdivided into equally-spaced points, and weights are computed for each of them. We empirically found an optimum value of 100µm for the distance among the points, where the position of the detected event arising from its energy levels (50-415 keV) into the detectors' array.Let p=n/N be the probability, from a geometrical point of view, that an event is detected, where n is the number of crystals inside the maximum lighted surface (Fig. 2, right, the shaded area on the left) and N is the total number of crystals, here 400. Then the weight w for that point is the inverse of the probability p. Hence, a point belonging to the solid angle defined by the dotted lined crossing the centre O has the maximum probability p = 1 and the minimum weight w = 1. In contrast, a point belonging to a LOR joining two crystals placed at the corner of detectors has the minimum probability p = 1/400 and the maximum weight w = 400. Here the kernel of the sequential algorithm is briefly outlined, in a pseudo-code: 1. for (each accepted event) 2. \ for (each crossed voxel) 3. \ \ for (x from 1 to l/r) 4. \ \ \ V(ix, jx, kx) = V(ix, jx, kx) + w 5. filter (V) where line 4. represents the backprojection and V is the volume to be reconstructed, l is the length of the LOR and r depends on the resolution of the object (here r = 100µm). Parallel algorithm. There are no data dependencies among frames, the BPF algorithm is intrinsically parallel and each event can be processed independently and the BPF reconstruction, accordingly. Once all data have been processed, then filtering takes place. To accomplish the parallel version of this algorithm we implement the SPMD (Single Program Multi Data) model required in which one code runs on each processor. We follow the master-slave paradigm, in which a separate master program is responsible for assigning data to the processes (slaves), and for collecting the results. In practice, the filter procedure has not been parallelized because it takes a very small percentage of the total time. For more details see [MI12]. RESULTS
For our experiments we used a hot resolution phantom filled with a solution of radioactive water containing 18F
labelled fluoro-deoxyglucose (FDG): images of glucose utilization by the brain in FDG studies can be used to indicate
differences in glucose metabolism in vivo. This type of phantom is frequently used to assess the quality of images
produced by PET scanners. The phantom used is a 38 mm diameter by 40 mm height cylinder, consisting of arrays of cylindrical
"line" sources with 1.5, 2.0, 2.5 and 3.0 mm diameter, and centre-to-centre spacing equal to twice the source diameter.
Practically speaking, it lays into a volume of 48×48×48 mm3 represented by a 3D matrix M of dimension N3,
where N depends on the resolution of the final reconstructed volume. Here we use N = 64 with a resolution of 750µm.
Each head of the tomograph was rotated on a radius of 76 mm by 180° and detected over 30×106 events.
Fig. 4: First row: unfiltered transaxial slices: from left to right, taken at 8 mm (1/5 of the height), 20 mm (central slice) and 32 mm (4/5 of the height) from the upper side of the phantom. Second row: same slices after filtering. Image reconstruction can be easily performed in real time with our algorithm implementing dynamical load balancing on a small entry-level cluster of 8 Intel PIV 1.7GHz, inter-connected in a 100Mbit Ethernet LAN. The resolution could be improved by decreasingCombined tomographic and emission methods(CT, SPECT)OUTLINE
In recent years, nuclear imaging techniques such as PET (Positon Emission Tomography)
and SPECT (Single-Photon Emission Computed Tomography) have found
widespread oncologic use because they can visualize the metabolic activity of a tumor
with high sensitivity and specificity.
Although conventional mammography is the most widespread technique in clinical practice, its limitations
in terms of specificity and sensitivity are well known. To improve the diagnostic information scintigraphic
techniques have been used, and in particular scintimammography with an Anger camera. The recent introduction of
dedicated imagers with small field of view (FOV) coupled to a moderate breast compression has allowed to
increase the sensitivity up to 80% for tumour size between 0.5 and 1 cm. An alternative technique has also been
proposed with the breast in a pendular prone position and a dedicated ring for its SPECT imaging. It has been
suggested that a dedicated CT (Computed Tomography) system could be of great benefit if combined with this SPECT modality.
Indeed, by integrating the morphological (CT) image with the functional (SPECT) one, the sensitivity for lesions
smaller than 1 cm could be further increased.
METHOD An integrated CT–SPECT system for breast cancer study is being investigated. A dedicated tomographic imager has been assembled which allows us to perform scintimammography and X-ray CT in the same geometrical conditions. The experimetal set-up is schematically shown in Fig. 1. The system, whose toroidal ring has a diameter of about 13 cm, has been designed and built in order to image patients in a pendular prone position (leaning breast). In the prototype, two SPECT detectors are orthogonally positioned to the X-ray source-to-CT detector axis and have been mounted on the same rotating gantry. The X-ray source produces a quasi-monochromatic beam tuned at 28.5 keV (ΔE/E≈4%), with a measured flux of 1.7×104 photons mm2 s-1 mA-1. Fig. 1: Schematic of the integrated CT-SPECT system. The SPECT detector (two identical heads) is based on a scintillator matrix made up of 8×8 match-like crystals, 2.5×2.5×5 mm3 of size. The dimensions of the FOV (Field Of View) in SPECT are 2 cm axially and 2 cm in diameter. The energy resolution is 23% (FWHM) at 140 keV. CT acquisition. CT profiles were recorded by rotating the system with an angular step of 2° from 0° to 18°. The X-ray tube conditions were 40 kV, 10 mA, and the exposure time was 500 ms for all acquisitions. The following data processing must be applied to the raw profiles to obtain the corrected sinogram: (1) dark and white field normalization (2) calculation of -ln(I/I0) to correct the sinogram (3) centering of sinogram where I and I0 are, respectively, the transmitted intensity and the intensity in the air outside the object. The raw and the corrected sinograms are shown in Fig. 2, left and right, respectively. More details are given in [MI13]. Fig. 2: CT acquisition of a cylindrical phantom: raw sinogram (left), corrected sinogram (right). (1) the spatial non-linearity of the PSPMT (Position Sensitive Photo-Multiplier Tubes), i.e., the spatial distortions due to PSPMT were corrected via the appropriate algorithm (see Fig. 3) (2) the uneven response of PSPMT, i.e., the 140 keV peak-shift of each detector element was corrected by applying a normalization procedure (3) the variation of detection efficiency, i.e., the number of events detected by each element was corrected by using the image of a uniform (flood) radioactive source As for CT acquisition, a similar procedure for determing the COR of the SPECT system was carried out before acquiring the data. Fig. 3: Correction of spatial distortion in SPECT: raw matrix (left), matrix corrected using the appropriate look-up table (right). As for tomographic image reconstruction, the most innovative contribution has been brought in the SPECT image reconstruction. Differently from the most used method where data previously acquired are back-projected onto image pixels, the method developed aims to minimize a cost function, where the cost is the minimum deviation between measured data and data as if they were generated by the ``image'' itself, also considering attenuation and scattering phenomena. This complex minimization problem has been solved by means of the parallel iterative Simulated Annealing (SA) algorithm we conceived and implemented on a parallel SMP (Symmetric Multi-Processor) system [MI14][MI15]. RESULTS
Fig. 4: Fusion of CT and SPECT images in diverse slices. Characterization of the skin cells and wrinkles by using capacitive image analysisOUTLINE
The automated analysis of images and sequences could play a crucial role in most of the computer vision systems
for medical images. The assessment of pathologies or treatments is often entirely still left to human (visual)
evaluation, because of the objective difficulty to develop automated (or computer aided) assessment systems,
also due to the human health being a critical issue. In the field of dermatology and dermocosmetics, providing
an automated and accurate characterization of the skin surface is a basic requirement to enable a quantitative
assessment of the effects of environmental exposure, physiological changes or dermocosmetic treatments. In normal
subjects, skin ageing is the most visible effect: with aging, the skin surface changes its appearance from both
physiological and morphological points of view. METHOD
We have developed the first approach known in scientific literature, and in industry, relying on a compact and low-cost
capacitive device to achieve automatically quantitative measurements in vivo and in real time.
Originally developed for fingerprint acquisition and recognition, the device can give a detailed map (12×18 mm2)
of the skin surface
in terms of grey level values with a resolution of 50µm/pixel (Fig. 1). A sample image of the ventral side of
the forearm can be compared with one achieved by means of a camera (Fig. 1). This is the body site commonly considered
in such studies because of its limited photoageing due to environmental factors [S17].
Differently from optical images, capacitive images are obtained by contact and are
insensitive to light changes. In general, the capacitive device outputs gray levels values darker under skin tissue than
under skin wrinkles. However, black spots may be given by skin hydration or pore sweat (Fig. 1, last two images on the right).
In addition, also acquisition pressures affect the darkness of the image: while the last two images of Fig. 1, on the right,
show a lack of pressure at the image borders, the first image on the left is acquired with a right pressure.
To conclude, the data acquisition stage being operator-dependent as well as the physiological factors, yield images presenting
a high unhomogeneity. Fig. 1: From left to right: the capacitive device and the capacitive image (31 y.o., male); just for comparison, an image of almost the same skin region achieved using a camera; example of "bad" images, with black spots yielded by skin hydration or pore sweat. In order to reduce the effects of the sensor noise and of the grey level deformation due to both pressure differences and the effects of skin hydration, we perform the normalization of a local contrast, after a flat field correction. Fig. 2, first row, right, shows three normalized samples where every pixels have same contrast, referring to a 1 year old (y.o.) baby, a 51 y.o. and a 94 y.o. females. are the original ones. As one can see, looking at the original samples (the first three images of Fig. 2), as a person gets on in years, more and more wrinkles appear to get deeper and deeper, and wider.We realized that the original images look like a DEM (Digital Elevation Map) whereby the value of a pixel represents an elevation rather than luminance intensity - the darker the pixel, the higher the elevation of the terrain point. In our images, cells represent mountains and wrinkles represent valleys. Therefore, we have thought that such images could be segmented by using the methods usually employed in geomorphologic applications. In particular, we focused our attention on watershed analysis which subdivides the image into low-gradient catchment basins (locally homogeneous connected sets of pixels) surrounded by high-gradient watershed lines (connected pixels exhibiting local maxima in gradient magnitude). Wrinkles are watershed lines (first three images of Fig. 2, second row) and cells (last three coloured images) are catchment basins [S18]. Since wrinkles are typically absorbed into their respective cells, the method is not suitable to detect wrinkles. Fig. 2: First row: From left to right: the capacitive images referring to a 1 year old (y.o.) baby, a 51 y.o. and a 94 y.o. females; normalized samples of the same images. Second row: From left to right: the normalized images with superimposed watershed lines; the coloured segmented cells. Several features have been extracted from the skin topographic structures by using different image processing methods, referring to geometrical, photometric or statistical properties, on the original as well as in the Wavelet transformed images [S19][S20]. Fig. 3, first row, shows on the left the histogram H(x) of the cells area (in pixel), while its cumulative histogram HC(x) is on the right (51 y.o., female). M is chosen so that P=(M, HC(M)) has derivative equal to 1. Then, the feature related to cells is f1=HC(M). Fig. 3, second row, shows how a feature related to wrinkles has been extracted. The original image (first) is normalized (second) and then wrinkles are enhanced using a bank of bidimensional Gaussian filters having 14 different orientations (third). The rightmost image shows the corresponding wrinkles orientation chart [S21]. On the enhanced image we have extracted the MLC (Mean Local Contrast). This measure is related just to the wrinkle component of each sample, thus expecting it is higher for aged people, and it is the second feature here discussed.Fig. 3: First row: the histogram H(x) of the cells area and the related cumulative histogram HC(x) (the 51 y.o. subject of Fig. 2). Second row: From left to right: original sample (44 y.o., male); the normalized image; the image with the wrinkles enhanced; the corresponding wrinkle orientation chart, from which another feature have been extracted. In order to investigate the capability of the capacitive device to permit accurate 2D and 3D geometrical measurements of the skin surface, we have developed a method to compare measurements of the same body site achieved by means of the capacitive device and of an optical profilometer, considered as being the reference approach, but not suitable for in-vivo studies, since it requires silicone skin replicas [S22]. Fig. 4, top left, shows a 3D profilometric 640×480 image, covering a field of view of 627.6×470.7 µm2, with a resolution 50 times smaller than capacitive one (0.981µm/pixel). Accordingly, we needed to employ our mosaicing techniques to build a profilometric mosaic (Fig. 4, top right), thus achieving a region wide enough to enable a comparison. Subsequently, we have conceived an accurate image fusion method to overlap the profilometric mosaic to the corresponding region of the capacitive image (Fig. 4, bottom left). The last step has been to develop a method that exploits B-spline to fit the skin surface and extracts absolute measures regarding the wrinkles' width and depth from both images. Fig. 4, bottom right, shows the native profilometric line profile and the reconstructed capacitive one. Control points in the saturated regions that are close to high gradients of the capacitive profile are automatically removed from the spline interpolation [S23].Fig. 4: First row: 3D profilometric map (left) and mosaic of profilometric images (right). Second row: fusion of the profilometric mosaic and the capacitive image (left); native profilometric line profile and the recostructed capacitive one. RESULTS
Fig. 5: First row: From left to right: correlation between capacitive and profilometric measurements of wrinkles' width and depth (in µm); results for features f1 and MLC. Skin surface analysis. After that the reliability of planar 2D measurements have been proved, we have assessed the reliability of our features by searching for the correlation with skin aging, that in normal subjects corresponds to biological aging and can be considered as the ground truth. A good correlation between features and skin ageing should validate the features as being reliable to characterize the skin surface, accordingly. Subsequently, a dataset of N=90 normal subjects has been built up together with a team of dermatologists during common dermatological examinations.Fig. 5, graphics 3 and 4, from left, report the results attained f1, after being normalized, and MLC. The good correlation between values of the features extracted (here, mainly f1) and skin aging proves how the features we conceived are suitable to measure the skin's health status. Besides, in order to assess the accuracy of our measurements for a given subject we have performed a measurement repeatability test on a set of 10 people, one for each decade. The relevant results achieved (e.g. f1 yields a relative error of 0.21%) prove that our system can be used to measure any change in the skin surface for follow-up studies in dermatology and dermocosmetics. At present, we have started a follow up study involving 30 subjects for evaluating middle/long term changes due to hydrating treatments. So far, preliminary experiments directed to assess the skin surface modifications due to hydrating creams [S25] as well as to hyperhydrosis have yielded promising results. It is worth remarking that we have been invited to present these results at the National Congress of the Italian Association of Hospital Dermatologists [S26]. |